Por favor, use este identificador para citar o enlazar este ítem:
http://repositoriodigital.ipn.mx/handle/123456789/11587| Título : | Sistema automatizado para la medicion de la conductividad termica de lıquidos mediante el metodo del alambre caliente |
| Palabras clave : | hot wire thermal conductivity |
| Fecha de publicación : | 16-ene-2013 |
| Editorial : | REVISTA MEXICANA DE FISICA |
| Descripción : | We present the implementation of an automated system based on the hot-wire technique for measurement of the thermal conductivity of
liquids using high precision equipment (current source and voltage meter) that allow to work without an array of resistances known as a
Wheatstone bridge, commonly used in such equipments. We also use a criterion to verify that we are working on the correct linear region of
the curve ¢T versus ln(t), which is used for experimental data processing. Finally we validate the functionality of the hot-wire experimental
array by measuring the thermal conductivity in samples of liquids with well-known thermal properties. INSTITUTO POLITECNICO NACIONAL, CONACYT |
| URI : | http://www.repositoriodigital.ipn.mx/handle/123456789/11587 |
| Otros identificadores : | http://hdl.handle.net/123456789/1124 |
| Aparece en las colecciones: | Doctorado |
Ficheros en este ítem:
| Fichero | Descripción | Tamaño | Formato | |
|---|---|---|---|---|
| RMF_Marin_2011_1.pdf | 37.17 kB | Adobe PDF | Visualizar/Abrir |
Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.